NPPTL Respirator Assessments to Support the COVID-19 Response
Updated April 16, 2021
As a consequence of the removal of these crisis capacity strategies, NPPTL will no longer conduct any of the Respirator Assessments to Support the COVID-19 Response.
Decontaminated Assessment Results
NPPTL has completed Decontaminated Assessments for the products listed below.
While decontamination and reuse of FFRs are not consistent with standard and approved usage, these options may need to be considered when FFR shortages exist. This assessment was developed to quantify the filtration efficiency and manikin fit factor of an N95 respirator that has been decontaminated. This assessment is not to determine the effectiveness of the decontamination procedure at killing pathogenic microorganisms. Respirators with exhalation valves are included in this assessment and are evaluated as received by NPPTL for manikin fit testing. Exhalation valves are sealed with wax for filtration testing and therefore, the performance of the exhalation valve is not examined for the filtration test. The results provided in this report are specific to the subset of samples that were provided to NPPTL for assessment. These results may be used to update the CDC guidance for Crisis Capacity Strategies (during known shortages).
Respirator Model | Decon. Method |
Decon. Cycles |
Filtration Efficiency Results | Strap Performance |
Manikin Fit Factor |
Test Report |
|
---|---|---|---|---|---|---|---|
Max. % | Min. % | ||||||
3M 1860 | ClorDiSys Solutions Minidox-M, Chlorine Dioxide Generator, Concentration: 1 mg/L, RH: 65%, Exposure: 720 ppm/hrs | 4 | 99.73 | 98.82 | No visual degradation > 15% change in force Discolored nosepiece foam |
≥100 | Report 3pdf icon |
3M 1860 | DiKlor-G® Sterilization | 6 | 99.62 | 99.13 | No visual degradation. >10% change in force Discolored nosepiece foam |
≥ 100 | Report 10pdf icon |
3M 8000* | Gas phase chlorine dioxide at 1000~2000 ppm·hr |
5 | Not assessed | Not assessed | Discolored straps <10% change in force |
≥ 100 | Report 11pdf icon |
3M 8000* | Gas phase chlorine dioxide at 1000~2000 ppm·hr |
10 | Not assessed | Not assessed | Discolored straps <15% change in force |
≥ 100 | Report 11pdf icon |
3M 8000 | Gas phase chlorine dioxide at 1000~2000 ppm·hr |
20 | 99.30 | 98.84 | Discolored straps <10% change in force |
Min: 92 Max: ≥100 |
Report 11pdf icon |
3M 8200* | Chlorine Dioxide at 2,500 ppm·hours | 4 | 99.59 | 99.39 | >10% change in force Discolored nosepiece foam and straps |
Not assessed | Report 31pdf icon |
3M 8200 | Chlorine dioxide gas, 5 x 850 ppm-hours | 5 | 99.52 | 98.74 | >10% change in force Discolored nosepiece foam and straps |
≥100 | Report 36pdf icon |
3M 8210 | DiKlor-G® Sterilization | 6 | 99.78 | 98.73 | No visual degradation >15% change in force Discolored nosepiece foam |
≥ 100 | Report 4pdf icon |
3M V-flex 1805 | ClorDiSys Solutions Minidox-M, Chlorine Dioxide Generator, Concentration: 1 mg/L, RH: 65%, Exposure: 720 ppm/hrs | 1 | 99.80 | 99.50 | No visual degradation <10% change in force |
≥100 | Report 3pdf icon |
VWR 89201 | ClorDiSys Solutions Minidox-M, Chlorine Dioxide Generator, Concentration: 1 mg/L, RH: 65%, Exposure: 720 ppm/hrs | 4 | 99.59 | 97.91 | No visual degradation <10% change in force |
Min: 37 Max: 96 |
Report 3pdf icon |
3M VFlex 1804* | Dense-phase CO2, 37°C above its supercritical pressure of 1100 psig with humidification for 90 min | 1 | 91.79 | 82.00 | Respirators were soiled and information blurred/illegible >10% change in force |
≥100 | Report 39pdf icon |
3M 1860 | Supercritical CO2 | 10 | 89.90 | 66.90 | Discolored respirator surface <5% change in force |
Min: 54 Max: ≥100 |
Report 20pdf icon |
3M 1860* | Dense-phase CO2, 37°C above its supercritical pressure of 1100 psig with humidification for 90 min | 1 | 74.70 | 55.70 | Respirators were soiled and information blurred/illegible >5% change in force |
Min: 22 Max: 42 |
Report 39pdf icon |
3M 1860S* | Dense-phase CO2, 37°C above its supercritical pressure of 1100 psig with humidification for 90 min | 1 | 84.90 | 57.70 | Respirators were soiled and information blurred/illegible >10% change in force |
Min: 8 Max: 13 |
Report 39pdf icon |
3M Vflex 1804 | Dry Heat, 75-78°C recirculating hot air for 45 minutes | 10 | 99.85 | 98.98 | Respirator information on front of respirator was blurred/illegible <5% change in force |
Min: 70 Max: ≥100 |
Report 37pdf icon |
3M VFlex 1804 | Dry Heat, 75-78 °C, 45 mins/cycle | 10 | 99.91 | 99.72 | Respirator information on front of respirator was blurred/illegible >10% change in force |
≥100 | Report 44pdf icon |
3M 1860 | Commercial laundry tumbler dryers, 70°C for 60 min |
3 | 99.33 | 98.53 | No visual degradation <5% change in force |
≥100 | Report 12pdf icon |
3M 1860 | Commercial laundry tumbler dryers, >80°C for >70 min |
3 | 98.89 | 95.08 | No visual degradation >5% change in force |
≥100 | Report 18pdf icon |
3M 1860* | Dry Heat , 75°C for 60 min per cycle | 5 | 99.02 | 98.32 | No visual degradation <5% change in force |
≥100 | Report 21pdf icon |
3M 1860 | Dry Heat, 75-78°C recirculating hot air for 45 minutes | 10 | 98.96 | 97.70 | Respirator information on front of respirator was blurred/illegible >5% change in force |
≥100 | Report 37pdf icon |
3M 1860 | MSU Extension Dry Heat Decon (85 °C for 50 minutes) | 4 | 99.02 | 98.14 | Discolored straps. Respirator information on front of respirator was blurred/illegible >15% change in force |
Min: 64 Max: ≥100 |
Report 40pdf icon |
3M 1860S | Dry Heat, 75-78°C recirculating hot air for 45 minutes | 10 | 98.79 | 97.86 | Respirator information on front of respirator was blurred/illegible >5% change in force |
≥100 | Report 37pdf icon |
3M 1870+* | Dry Heat , 75°C for 60 min per cycle | 5 | 99.81 | 98.60 | No visual degradation <5% change in force |
≥100 | Report 21pdf icon |
3M Aura 1870+ | Dry Heat, 75-78°C recirculating hot air for 45 minutes | 10 | 99.98 | 98.95 | <5% change in force | ≥100 | Report 37pdf icon |
3M 8210 | MSU Extension Dry Heat Decon (85 °C for 50 minutes) | 4 | 99.96 | 97.33 | Respirator information on front of respirator was blurred/illegible >40% change in force |
≥100 | Report 40pdf icon |
3M 8210+ | Dry heat, 75°C for 30 min | 10 | 99.45 | 98.66 | No visual degradation <5% change in force |
≥100 | Report 25pdf icon |
3M 8511 | Commercial laundry tumbler dryers, 70°C for 60 min |
3 | 99.82 | 99.09 | No visual degradation <5% change in force |
≥100 | Report 12pdf icon |
3M 8511 | Commercial laundry tumbler dryers, >80°C for >70 min |
3 | 99.12 | 98.14 | No visual degradation >10% change in force |
≥100 | Report 18pdf icon |
3M Aura 9205+ | Dry Heat, 75-78°C recirculating hot air for 45 minutes | 10 | 99.91 | 99.39 | >5% change in force | ≥100 | Report 37pdf icon |
Halyard 46727* | Dry Heat , 75°C for 60 min per cycle | 5 | 99.54 | 98.76 | No visual degradation <5% change in force |
≥100 | Report 21pdf icon |
Halyard 62126 | Commercial laundry tumbler dryers, 70°C for 60 min |
3 | 99.78 | 99.18 | No visual degradation <5% change in force |
≥100 | Report 12pdf icon |
Halyard 62126 | Commercial laundry tumbler dryers, >80°C for >70 min |
3 | 99.32 | 97.31 | No visual degradation >5% change in force |
≥100 | Report 18pdf icon |
Halyard Fluidshield 46767 Regular | The Clean Sleep Hyperion, 90 mins at 175°F | 5 | 93.77 | 61.90 | <5% change in force Respirators were significantly damaged. |
Min: 2 Max: 5 |
Report 32pdf icon |
Halyard Fluidshield 46767 Regular | Clean Sleep Hyperion, 90 mins at 175°F | 3 | 99.62 | 97.89 | No visual degradation <5% change in force |
≥100 | Report 35pdf icon |
3M 1870* | Gaseous ozone, 450 ppm | 5 | 99.93 | 99.72 | No visual degradation >5% change in force |
≥100 | Report 6pdf icon |
3M 1860* | Gravity Steam Cycle | 3 | 98.24 | 96.93 | No visual degradation >5% change in force |
Min: 18 Max: 20 |
Report 15pdf icon |
3M 1860* | Cres Cor moist heat decontamination system: 1 hour of exposure at 100°C and 100% RH |
1 | 98.02 | 97.38 | No visual degradation. >5% change in force. Respirator information on front of respirator was blurred/illegible. |
Min: 20 Max: 92 |
Report 26pdf icon |
3M 1860* | Cres Cor moist heat decontamination system: 1 hour of exposure at 100°C and 100% RH |
3 | 98.45 | 97.24 | No visual degradation of the straps. >10% change in force. Respirator information on front of respirator was blurred/illegible. |
Min: 15 Max: 38 |
Report 26pdf icon |
3M 1860* | Cres Cor moist heat decontamination system: 1 hour of exposure at 100°C and 100% RH |
5 | 98.08 | 96.92 | No visual degradation of the straps. >10% change in force. Respirator information on front of respirator was blurred/illegible. |
Min: 24 Max: 51 |
Report 26pdf icon |
3M 1860* | Cres Cor moist heat decontamination cabinet, 15 mins at 85°C and 65% RH |
1 | 99.34 | 99.08 | <5% change in force Blurred print on front of respirators |
Min: 82 Max: 114 |
Report 33pdf icon |
3M 1860* | Cres Cor moist heat decontamination cabinet, 15 mins at 85°C and 65% RH |
3 | 99.25 | 99.05 | <5% change in force Blurred print on front of respirators |
Min: 86 Max: 108 |
Report 33pdf icon |
3M 1860* | Cres Cor moist heat decontamination cabinet, 15 mins at 85°C and 65% RH |
5 | 98.92 | 98.04 | <5% change in force Blurred print on front of respirators |
Min: 93 Max: 116 |
Report 33pdf icon |
3M 1870* | Gravity Steam Cycle | 3 | 98.59 | 97.49 | No visual degradation >20% change in force |
≥100 | Report 15pdf icon |
3M 8210 | Electric Steamer, 100°C for 30 sec/cycle | 10 | 99.46 | 98.03 | Respirator information on front of respirator was blurred/illegible >10% change in force |
Min: 18 Max: 22 |
Report 38pdf icon |
3M V-flex 9105* | Gravity Steam Cycle | 3 | 98.99 | 98.47 | No visual degradation >10% change in force |
≥100 | Report 15pdf icon |
3M 8200* | High energy (10 MeV), high dose rate (3 kGy/s) electron beam irradiation at 25 kGy per decontamination cycle. | 1 | 55.7 | 46.6 | No visual degradation <5% change in force |
Not assessed | Report 5pdf icon |
3M 8200* | High energy (10 MeV), high dose rate (3 kGy/s) electron beam irradiation at 25 kGy per decontamination cycle. | 2 | 51.1 | 47.2 | No visual degradation >10 % change in force |
Not assessed | Report 5pdf icon |
3M 8200* | High energy (10 MeV), high dose rate (3 kGy/s) electron beam irradiation at 25 kGy per decontamination cycle. | 3 | 50.6 | 47.4 | No visual degradation <15% change in force |
Not assessed | Report 5pdf icon |
3M 8200* | Electron Beam Irradiation | 1 | 63.6 | – | No visual degradation. >10% change in force |
Not assessed | Report 27pdf icon |
N95 3M 1860S | Methylene Blue | 5 | 98.65 | 93.59 | No visual degradation <5% change in force |
≥100 | Report 1pdf icon |
3M 1860* |
Methylene Blue + Light |
5 | 99.00 | 98.34 | No visual degradation <5% change in force |
≥100 | Report 21pdf icon |
3M 1870+* |
Methylene Blue + Light |
5 | 99.88 | 99.64 | No visual degradation <10% change in force |
≥100 | Report 21pdf icon |
Halyard 46727* |
Methylene Blue + Light |
5 | 99.44 | 97.24 | No visual degradation <5% change in force |
≥100 | Report 21pdf icon |
3M 1860* | Microwave generated plasma | 3 | 99.50 | 98.81 | No visual degradation >5% change in force |
≥100 | Report 24pdf icon |
3M 8210* | Microwave generated plasma | 3 | 99.33 | 91.54 | No visual degradation >5% change in force |
Min: 83 Max: ≥100 |
Report 19pdf icon |
3M 8511 | Sterrad NX100 Hydrogen Peroxide Vapor/Low Temp plasma |
2 | 78.90 | 58.30 | No visual degradation <5% change in force |
Min: 40 Max: ≥100 |
Report 16pdf icon |
Halyard Fluidshield 46727 Reg | Sterrad NX100 Hydrogen Peroxide Vapor/Low Temp plasma |
2 | 98.82 | 98.29 | No visual degradation >5% change in force |
≥100 | Report 8pdf icon |
3M 1860* | UV-Concepts Inc. model UVE (UV-C Enclosure). 180 seconds and produces ~1000 mJ/cm² of UV-C “dose.” | 10 | 99.61 | 98.97 | No visual degradation <5% change in force. |
>100 | Report 30pdf icon |
3M 1860S* | Broadband UVC (DECON95) | 1 | 99.24 | 98.58 | No visual degradation. <5% change in force |
Not assessed | Report 29pdf icon |
3M 1860S* | Broadband UVC (DECON95) | 3 | 99.22 | 97.96 | No visual degradation. <5% change in force |
Not assessed | Report 29pdf icon |
3M 1860S* | Broadband UVC (DECON95) | 5 | 98.89 | 98.12 | No visual degradation. >5% change in force |
Not assessed | Report 29pdf icon |
3M 8210* | UV-C 254 nm – 500 mJ/cm2 |
10 | 99.61 | 99.15 | No visual degradation >5% change in force |
≥100 | Report 22pdf icon |
3M 8210* | UV-C 254 nm – 1000 mJ/cm2 |
10 | 99.41 | 99.10 | No visual degradation >5% change in force |
≥100 | Report 22pdf icon |
3M 8210 | UV-C dose ˃ 2 J/cm² for 1 min/cycle | 20 | 99.70 | 98.37 | No visual degradation >10% change in force |
≥100 | Report 42pdf icon |
Moldex 1513 Large | UV-C dose ˃ 2 J/cm² for 1 min/cycle | 20 | 99.97 | 98.38 | Separated nosepiece foam on 1 sample <5% change in force | Min: 16 Max: ≥100 |
Report 41pdf icon |
3M 1860* | 1.06% peracetic acid (PAA) solution | 1 | 99.02 | 98.58 | No visual degradation. <5% change in force |
>100 | Report 28pdf icon |
3M 1860* | 1.06% peracetic acid (PAA) solution | 3 | 99.49 | 99.08 | No visual degradation. <5% change in force |
>100 | Report 28pdf icon |
3M 1860* | 1.06% peracetic acid (PAA) solution | 5 | 99.29 | 98.64 | No visual degradation. <5% change in force |
>100 | Report 28pdf icon |
BYD DE2322* | Plasma Discharge ROS | 1 | 98.04 | 97.79 | No visual degradation. <5% change in force |
Not assessed | Report 27pdf icon |
BYD DE2322* | Plasma Discharge ROS | 3 | 98.69 | 97.10 | No visual degradation. <5% change in force |
Not assessed | Report 27pdf icon |
Prestige Ameritech RP88020* | Plasma Discharge ROS | 1 | 99.56 | 99.35 | No visual degradation. <5% change in force |
Not assessed | Report 27pdf icon |
Prestige Ameritech RP88020* | Plasma Discharge ROS | 3 | 99.59 | 99.50 | No visual degradation. <5% change in force |
Not assessed | Report 27pdf icon |
3M 1860* | VPHP | 5 | 99.54 | 99.26 | No control provided | ≥100 | Report 2pdf icon |
3M 1860 | VPHP | 10 | 99.47 | 99.02 | No visual degradation ≤ 35% change in force |
≥100 | Report 7pdf icon |
3M 1860 | VPHP | 15 | 99.75 | 99.32 | No visual degradation >40% change in force |
≥100 | Report 14pdf icon |
3M 1860 | VPHP | 20 | 99.74 | 99.16 | No visual degradation >35% change in force |
≥100 | Report 14pdf icon |
3M 1860 | Bioquell HPV, >480 ppm HPV at 900 sec/cycle | 5 | 99.37 | 98.67 | No visual degradation >15% change in force |
≥100 | Report 43pdf icon |
3M 1860* | VHP | 5 | 98.97 | 97.67 | Discolored nosepiece foam >30% change in force |
≥100 | Report 21pdf icon |
3M 1860S* | VPHP | 5 | 99.73 | 99.01 | No visual degradation < 15% change in force |
≥100 | Report 2pdf icon |
3M 1870+* | VPHP | 20 | 99.78 | 98.00 | No visual degradation >15% change in force |
≥100 | Report 17pdf icon |
3M 1870+* | VHP | 5 | 99.89 | 99.52 | No visual degradation >10% change in force |
≥100 | Report 21pdf icon |
3M 8000* | VHP 7% | 1 | 98.99 | 98.69 | No control provided | Not assessed | Report 9pdf icon |
3M 8000* | VHP 7% | 2 | 99.05 | 98.33 | No control provided | Not assessed | Report 9pdf icon |
3M 8000* | VHP 7% | 3 | 98.80 | 98.00 | No control provided | Not assessed | Report 9pdf icon |
3M 8000* | VHP 7% | 4 | 99.18 | 98.74 | No control provided | Not assessed | Report 9pdf icon |
3M 8000* | VHP 7% | 5 | 99.16 | 98.78 | No control provided | Not assessed | Report 9pdf icon |
3M 8210* | VPHP | 5 | 99.72 | 99.48 | No visual degradation > 10% change in force |
≥100 | Report 2pdf icon |
3M 8210 | VPHP | 10 | 99.60 | 99.37 | No visual degradation ≤ 10% change in force |
≥100 | Report 7pdf icon |
3M 8210* | VHP 7% | 1 | 99.53 | 97.26 | No control provided | ≥100 | Report 9pdf icon |
3M 8210* | VHP 7% | 2 | 99.41 | 99.36 | No control provided | ≥100 | Report 9pdf icon |
3M 8210* | VHP 7% | 3 | 99.50 | 97.60 | No control provided | ≥100 | Report 9pdf icon |
3M 8210* | VHP 7% | 4 | 99.52 | 99.31 | No control provided | ≥100 | Report 9pdf icon |
3M 8210* | VHP 7% | 5 | 99.49 | 99.09 | No control provided | ≥100 | Report 9pdf icon |
3M 8210 | VPHP | 20 | 99.90 | 98.19 | No visual degradation >15% change in force |
≥100 | Report 14pdf icon |
3M 8210V* | VHP 7% | 1 | 98.94 | 98.24 | No control provided | ≥100 | Report 9pdf icon |
3M 8210V* | VHP 7% | 2 | 98.79 | 98.57 | No control provided | ≥100 | Report 9pdf icon |
3M 8210V* | VHP 7% | 3 | 98.53 | 98.30 | No control provided | ≥100 | Report 9pdf icon |
3M 8210V* | VHP 7% | 4 | 99.10 | 97.97 | No control provided | ≥100 | Report 9pdf icon |
3M 8210V* | VHP 7% | 5 | 98.62 | 98.34 | No control provided | ≥100 | Report 9pdf icon |
3M 8511 | VPHP | 5 | 99.81 | 99.63 | No visual degradation >20% change in force |
≥100 | Report 14pdf icon |
3M 8511 | VPHP | 10 | 99.88 | 99.46 | No visual degradation >20% change in force |
≥100 | Report 14pdf icon |
3M 8511 | VPHP | 20 | 99.73 | 99.26 | No visual degradation >15% change in force |
≥100 | Report 13pdf icon |
3M 9205+ | VPHP | 5 | 99.86 | 98.11 | No visual degradation <15% change in force |
≥100 | Report 23pdf icon |
3M 9205+ | VPHP | 10 | 99.78 | 97.77 | No visual degradation <15% change in force |
≥100 | Report 23pdf icon |
3M 9205+ | VPHP | 20 | 99.69 | 96.13 | No visual degradation <15% change in force |
≥100 | Report 23pdf icon |
3M V-flex 1804* | VPHP | 5 | 99.71 | 99.50 | No visual degradation >5% change in force |
≥100 | Report 2pdf icon |
3M V-flex 1804 | VPHP | 10 | 99.59 | 99.28 | No visual degradation ≤ 25% change in force |
Min: 27 Max: ≥100 |
Report 7pdf icon |
3M V-Flex 1804 | VPHP | 20 | 99.67 | 99.00 | Discolored straps >5% change in force |
Min: 85 Max: ≥100 |
Report 14pdf icon |
Crosstex GPRN95* | VHP 7% | 1 | 99.66 | 99.09 | No control provided | Not assessed | Report 9pdf icon |
Crosstex GPRN95* | VHP 7% | 2 | 99.62 | 99.03 | No control provided | Not assessed | Report 9pdf icon |
Crosstex GPRN95* | VHP 7% | 3 | 99.28 | 97.85 | No control provided | Not assessed | Report 9pdf icon |
Crosstex GPRN95* | VHP 7% | 4 | 99.22 | 96.78 | No control provided | Not assessed | Report 9pdf icon |
Crosstex GPRN95* | VHP 7% | 5 | 99.13 | 96.84 | No control provided | Not assessed | Report 9pdf icon |
Gerson 1730 | VPHP | 5 | 99.65 | 99.26 | No visual degradation. <5% change in force |
≥100 | Report 17pdf icon |
Gerson 1730 | VPHP | 10 | 99.44 | 97.77 | No visual degradation >15% change in force |
≥100 | Report 17pdf icon |
Gerson 1730 | VPHP | 20 | 99.48 | 99.20 | No visual degradation >5% change in force |
≥100 | Report 17pdf icon |
Gerson 1740 | VPHP | 5 | 99.35 | 98.82 | No visual degradation >10% change in force |
Min: 88 Max: ≥100 |
Report 17pdf icon |
Gerson 1740 | VPHP | 10 | 99.40 | 98.47 | No visual degradation >5% change in force |
Min: 68 Max: ≥100 |
Report 17pdf icon |
Gerson 1740 | VPHP | 20 | 99.36 | 98.58 | No visual degradation >5% change in force |
≥100 | Report 17pdf icon |
Halyard 46727* | VHP | 5 | 99.33 | 97.89 | No visual degradation >15% change in force |
≥100 | Report 21pdf icon |
Halyard 46727 | Bioquell HPV, >480 ppm HPV at 900 sec/cycle | 5 | 99.74 | 98.44 | No visual degradation <5% change in force |
≥100 | Report 43pdf icon |
Makrite 9500-N95S* | H₂O₂ (35% concentration) + catalyst | 10 | 99.46 | 96.15 | >5% change in force Discolored nosepiece foam. |
≥100 | Report 34pdf icon |
Makrite 9500-N95S* | H₂O₂ (35% concentration) + catalyst | 20 | 99.55 | 85.60 | <5% change in force Discolored nosepiece foam |
Min: 44 Max: ≥100 |
Report 34pdf icon |
Makrite 9500-N95S* | H₂O₂ (35% concentration) + catalyst | 30 | 99.54 | 98.10 | >10% change in force Discolored nosepiece foam |
Min: 28 Max: 88 |
Report 34pdf icon |
Moldex 1512* | VHP 7% | 1 | 98.79 | 97.91 | No control provided | ≥100 | Report 9pdf icon |
Moldex 1512* | VHP 7% | 2 | 99.05 | 98.74 | No control provided | ≥100 | Report 9pdf icon |
Moldex 1512* | VHP 7% | 3 | 98.59 | 97.80 | No control provided | ≥100 | Report 9pdf icon |
Moldex 1512* | VHP 7% | 4 | 98.78 | 97.76 | No control provided | ≥100 | Report 9pdf icon |
Moldex 1512* | VHP 7% | 5 | 98.74 | 97.55 | No control provided | ≥100 | Report 9pdf icon |
Moldex 1512 | VPHP | 20 | 99.53 | 97.56 | No visual degradation >15% change in force |
≥100 | Report 14pdf icon |
Moldex 1512 Medium* | VPHP | 5 | 99.12 | 96.25 | No visual degradation > 10% change in force |
≥100 | Report 2pdf icon |
Moldex 1512 Medium | VPHP | 10 | 99.40 | 95.19 | No visual degradation ≤ 10% change in force |
≥100 | Report 7pdf icon |
Moldex 2200 | VPHP | 20 | 99.46 | 99.38 | No visual degradation <5% change in force |
Min: 90 Max: ≥100 |
Report 14pdf icon |
Moldex 2200 M/L* | VPHP | 5 | 99.25 | 97.54 | No visual degradation ≤ 5% change in force |
Min: 34 Max: ≥100 |
Report 2pdf icon |
Moldex 2200 M/L | VPHP | 10 | 98.76 | 94.27 | No visual degradation ≤ 10% change in force |
≥100 | Report 7pdf icon |
Sperian N1105 | VPHP | 10 | 99.88 | 99.74 | No visual degradation >10% change in force |
≥100 | Report 13pdf icon |
Sperian N1105 | VPHP | 20 | 99.94 | 99.75 | No visual degradation >10% change in force |
Min: 58 Max: ≥100 |
Report 14pdf icon |
Sperian N1105 M/L | VPHP | 5 | 99.85 | 99.54 | No visual degradation ≤ 20% change in force |
≥100 | Report 7pdf icon |
Sperian N1125 | VPHP | 10 | 99.83 | 99.24 | No visual degradation >5% change in force |
Min: 89 Max: ≥100 |
Report 14pdf icon |
Sperian N1125 | VPHP | 20 | 99.85 | 99.33 | No visual degradation >5% change in force |
Min: 58 Max: ≥100 |
Report 14pdf icon |
Sperian N1125 M/L | VPHP | 5 | 99.88 | 99.13 | No visual degradation ≤ 15% change in force |
Min: 99 Max: ≥100 |
Report 7pdf icon |
Sperian One-Fit | VPHP | 5 | 99.75 | 98.90 | No visual degradation ≤ 5% change in force |
Not assessed | Report 7pdf icon |
Sperian One-fit | VPHP | 10 | 99.80 | 97.01 | No visual degradation >15% change in force |
Not assessed | Report 14pdf icon |
Sperian One-fit | VPHP | 20 | 99.85 | 99.42 | No visual degradation >15% change in force |
Not assessed | Report 14pdf icon |
* Testing for this model did not meet the requirements of the Decontaminated Respirator Assessment Plan because an insufficient number of samples were provided to NIOSH for 1 or more tests. Therefore, results are posted for the number of samples available. Please see results report for actual number of samples evaluated for filtration efficiency, manikin fit, and strap performance.