NIOSH logo and tagline

Two Fire Fighters Injured When an Automated External Defibrillator Battery Exploded or Ruptured During Testing – Florida

 

FF ShieldDeath in the Line of Duty…A summary of a NIOSH fire fighter fatality investigation

F99-39 Date Released: April 10, 2000

SUMMARY

On October 15, 1999, at 0700 hours, the Apparatus Operator on Engine 4 was conducting a routine equipment check when he experienced problems with the Medtronic Physio-Control LIFEPAK® 500, which is an automated external defibrillator (AED). He brought the malfunctioning AED to the attention of the Assistant Chief and a Captain, explaining that when he turned on the AED, it shut down instead of self-testing. The Assistant Chief and a Captain had the AED brought into one of the fire department offices, where they placed it on the desk to check it out. When the AED was turned on again, the units message system indicated the battery should be replaced. They removed and checked the lithium battery and noted it had an expiration date of 2004. They replaced the battery in the unit and it was turned on. An explosion or battery rupture occurred, which blew out the back of the AED (see photos).

It was reported by the Assistant Chief and the Captain that the concussion from the explosion or battery rupture blew the Assistant Chief out the office door onto the floor and blew the Captain against the office wall. The office was filled with white acrid smoke which made breathing difficult. The Assistant Chief reported he received flash burns to the left arm, a sprained left wrist, and an injured back. The Captain reported he suffered upper respiratory inflammation from the acrid smoke. NIOSH investigators concluded that to reduce the risk of similar occurrences fire departments, emergency medical services, and other users of AEDs should

  • follow the manufacturers instructions to replace battery packs immediately when the unit indicates a low battery or replace battery message

Additionally, NIOSH recommends

  • the Food and Drug Administration, which is responsible for approval of medical devices, should continue follow up with Medtronic Physio-Control, manufacturer of the LIFEPAK® AED, to reduce the potential hazards with battery packs in these types of devices

INTRODUCTION

On October 15, 1999, the Assistant Chief and a Captain were attempting to test a Medtronic Physio-Control LIFEPAK® 500 when it shut off automatically instead of performing a self test. Further investigation into the malfunction led to an eventual explosion or battery rupture of the battery pack, and it was reported by the Assistant Chief and the Captain that the concussion of the explosion or battery rupture blew the Assistant Chief out the office door onto the floor and blew the Captain against the office wall.

On October 20, 1999, the International Association of Fire Fighters (IAFF) requested that NIOSH provide technical assistance in reviewing the circumstances surrounding this incident. On October 27, 1999, an investigation was conducted by the NIOSH Senior Fire Fighter Investigator and an Investigator with the Food and Drug Administration. [Note: External defibrillators are medical devices which come under the jurisdiction of the Food and Drug Administration.] An opening conference was conducted with the Chief, the Assistant Chief (victim #1), and the Captain (victim #2) of the fire department and with the President of the local IAFF. The following persons were interviewed: the Chief, the Assistant Chief (victim #1), and the Captain (victim #2) of the fire department and representatives of the State Fire Marshal’s office. Since the AED involved in this incident had been picked up and a replacement provided by the manufacturer, Medtronic Physio-Control, the replacement AED was photographed and examined. The fire department provided a product safety advisory bulletin issued by the State (see Attachment #1), a copy of the customer letter from the manufacturer (see Attachment #2), a copy of the State Fire Marshal’s report, and written fire fighter statements. A video taken of the aftermath of the explosion was also provided by the fire department, and several still photographs were also reviewed. Several telephone and e-mail interviews were conducted with the Vice President of Quality and Regulatory Affairs with Medtronic Physio-Control, the manufacturer of the AED, to check on the status of the internal investigation being conducted by the manufacturer. The attachments to this report were provided by Medtronic Physio-Control.

The fire department involved in this incident serves a population of 6,000 in a geographic area of 3 square miles. The fire department operates one fire station and is comprised of 16 uniformed fire fighters. The fire department requires all new fire fighters to complete the State requirements for fire fighters at the fire training academy. The State requirements are equivalent to the requirement of the National Fire Protection Association Level I.

INVESTIGATION

On October 15, 1999, at the start of the 0700 hours shift, a fire fighter was conducting a check of the equipment on Engine 4. The fire fighter pulled the LIFEPAK® 500 automated external defibrillator (AED) from the engine, took it inside, and placed it on a table to perform a routine check. The fire fighter turned on the unit, and the Light Emitting Diode (LED) readout flashed on then shut off. He took the unit back to the engine and attempted to turn it on, however, nothing happened. The fire fighter informed the Assistant Chief of the problem and demonstrated what had occurred. He turned the unit on, and the LED readout indicated the battery should be replaced. The Assistant Chief and a Captain took the AED into a small office to attempt to determine the problem. The AED was placed on a desk, and the Assistant Chief pulled the lithium battery pack to check the expiration date, which was 2004. The battery pack was replaced in the AED, and when the unit was turned on, an explosion or battery rupture occurred. The explosion or battery rupture blew out the back of the AED (see photo #2), and it was reported by the Assistant Chief and Captain that the concussion blew the Assistant Chief out the office door onto the floor and blew the Captain against the office wall. The office was filled with white acrid smoke which made breathing difficult. The Assistant Chief reported that he received flash burns to the left arm, a sprained left wrist, and an injured back, and the Captain reported that he suffered upper respiratory inflammation from the acrid smoke. Both fire fighters were transported to a local hospital, treated, and released.

According to the State Fire Marshal’s report and interviews that were conducted, the lithium battery pack, which is located in the rear of the unit, either exploded or ruptured. The explosion or rupture blew the rear of the unit apart and sent pieces flying throughout the room. The explosion or rupture projected the AED in an upward trajectory.

Three other documented incidents involving this company’s AED units have occurred. According to the manufacturer, all three incidents were unwitnessed, and the explosion or rupture was contained by the soft carrying case. No injuries were reported in any of these incidents.

The manufacturer of this unit, Medtronic Physio-Control, has reported there have been 4 battery pack failures in 30,000 battery packs that are in use over the last 3 years.

RECOMMENDATIONS/DISCUSSION

Recommendation #1: Fire departments, emergency medical services, and other users of AEDs should follow the manufacturers instructions to replace battery packs immediately when the unit indicates a low battery or replace battery message.

Discussion: The manufacturer recommends that when a unit indicates a low battery or replace battery warning, then the battery has reached depletion due to normal use or even abnormal conditions, and the battery should be replaced. It is recommended that the operator or user of the unit not attempt to trouble shoot the battery warning by repeated attempts of reinserting the old battery. It is recommended that a spare battery pack be readily available for replacement.

Recommendation #2: The Food and Drug Administration, which is responsible for approval of medical devices, should continue follow up with Medtronic Physio-Control, manufacturer of the LIFEPAK® AED, to reduce the potential hazards with battery packs in these types of devices.

REFERENCES

National Institute for Occupational Safety and Health (NIOSH). 1997. Fact Sheet, Exploding Flashlights. NIOSH Publication No. 97-149.

INVESTIGATOR INFORMATION

This investigation was conducted by Ted A. Pettit, Senior Fire Fighter Investigator, NIOSH, Division of Safety Research, and Brant M. Schroeder, Consumer Safety Officer, Food and Drug Administration.

AED with Battery

Photo 1. AED with Battery

AED after Explosion or Rupture

Photo 2. AED after Explosion or Rupture

Battery Pack after Explosion or Rupture

Photo 3. Battery Pack after Explosion or Rupture

ATTACHMENT #1

DATE: October 20, 1999
TO: Fire Chiefs, Law Enforcement and Emergency Services Personnel
FROM: State of Florida, Division of State Fire Marshal
SUBJECT: Update on LIFEPAK® 500 AED

This notice is to serve as an update to this Division’s previous safety advisory dealing with explosions in the above named product manufactured by Medtronic Physio-Control Corporation.

The State fire Marshal’s Office has been in contact with numerous agencies in an attempt to develop all information possible to relay it to the appropriate agencies. During this investigation it has come to our attention that a possible “panic” condition is developing due to information spreading from unconfirmed and possible unreliable sources (i.e., the Internet).

The State Fire Marshal’s Office has not as of yet made any determinations of recommendations as to the continued use of the above products within the State of Florida. We are currently awaiting the results of testing being conducted by Medtronic Physio-Control Corporation and their related suppliers. Medtronic Physio-Control has been co-operating with the State Fire Marshals Office in this investigation and has pledged their continued co-operation. They advised that the incidence of any problems with their products worldwide have been extremely low but understand that any problems like those previously mentioned are cause for concern where lives & safety are at stake. They have asked that we include the attached letter from Medtronic Physio-Control Corporation with this update. (see Attachment #2)

At this time the exact cause for these incidents is still being investigated. We presently feel that Medtronic Physic-Control Corporation is interested in determining the source of these problems and address the needs for its correction.

All fire service, law enforcement personnel and other users of AED equipment should still be cognizant of the potential safety hazard but temper that with the fact that these devices are still an important piece of lifesaving equipment. At this time any discontinuance of use must be an agency decision.

If you have any questions or concerns you may contact Mr. Mitch Parrish, Vice President of Service, Medtronic Physio-Control Corporation at (425) 867-4491.

ATTACHMENT #2

October 19, 1999

Dear Customer,

This letter is to provide you with accurate and up-to-date information about an incident involving the lithium batteries used in some LIFEPAK® 500 automated external defibrillators. We are concerned that some press and e-mail messages being circulated are not only incorrect but may cause unnecessary concern.

Since the LIFEPAK® 500 AED was introduced two and one-half years ago there have been four documented cases of sudden venting of a lithium battery, none involving patient applications. In one case reported to us on Friday, October 15, the battery was being handled by a firefighter and caused a minor injury; he and a co-worker were treated and discharged from a local emergency room.

All batteries, whether used in a medical device, your cell phone, a flashlight or any other application, are chemical packages that generate and store electrical energy, and have the potential to vent or rupture. The incidence of this problem is rare. We have had four occurrences in a population of approximately 144,000 lithium cells used in Physio products, some of which have been in service more than two years. Worldwide, there are another 360,000 cells in use in other applications; we have been advised of one other similar incident in this additional population.

We, in conjunction with the battery manufacturer, are conducting an evaluation of the four cells which have vented to ensure the venting was not due to an avoidable circumstance. From our initial review there is no indication to date of such a cause. We are continuing our evaluation and intend to exhaust all possibilities.

Our commitment is to provide lifesaving tools to prevent unnecessary death due to sudden cardiac arrest. We are also committed to protect our customers’ safety without reducing the positive effect of having these lifesaving devices in places where sudden cardiac arrest may occur.

We remain convinced this battery technology is appropriate for use in AEDs. We recommend that you keep your devices in service.

We stand behind the quality of our products. If, at any time, you have questions about this issue or concerns about any Medtronic Physio-Control product, please call our customer hotline at 1.800.732.3088.

Sincerely,

MEDTRONIC PHYSIO-CONTROL CORP.
Richard O. Martin, Ph.D., P.E.
President

This page was last updated on 04/13/00