NIOSH
HAZARD REVIEW
Health
Effects of Occupational Exposure
to
Respirable Crystalline Silica
|
| |
|
| |
<Back |
|
|
 |
 |
 |
 |
| |
Table
8
|
| |
|
Table
8. Intralaboratory results for evaluation of XRD silica method
|
|
Item
|
Filter
loading
|
|
69.4
µg
|
98.4
µg
|
204
µg
|
| Degrees
of freedom |
12
|
11
|
12
|
| RSD
for sampling and analytical methods (%)*, |
8.8
|
6.3
|
8.1
|
Source:
NIOSH, BOM [1983].
*RSD = relative standard deviation. RSD for sampling and
analytical methods represents the RSD in mass estimates, accounting
for intersampler and analytical variability.
Implications for XRD: Pooled filter levels and pump
error (assumed to be <5%) indicate that the overall imprecision
is as follows: Total RSD for sampling and analytical methods is 9.3%.
Therefore, the upper 95% confidence limit on the accuracy (35 degrees
of freedom) is 21%. |
|
 |
 |
 |
 |
| |
Table
9
|
| |
|
Table
9. Intralaboratory results for evaluation of IR silica method
|
|
Item
|
Filter
loading
|
|
67.2
µg
|
99.7
µg
|
161µg
|
| Degrees
of freedom |
10
|
12
|
11
|
| RSD
for sampling and analytical methods (%)*, |
5.8
|
7.8
|
7.4
|
Source:
NIOSH, BOM [1983].
*RSD = relative standard deviation. RSD for sampling and
analytical methods represents the RSD in mass estimates, accounting
for intersampler and analytical variability.
Implications for IR: Pooled filter levels and pump
error (assumed to be <5%) indicate that the overall imprecision
is as follows: Total RSD for sampling and analytical methods is 7.1%.
Therefore, the upper 95% confidence limit on the accuracy (33 degrees
of freedom) is 17%. |
|
 |
 |
 |
 |
| |
Table
10
|
| |
|
Table
10. XRD method evaluation: concentration ranges bracketing applicable
exposure limits for which the NIOSH accuracy criterion is met*
( µg/m3)
|
|
Cyclone
and sampling rate
|
Filter
loading
|
Applicable
exposure limit
|
|
69.4
µg
|
98.4
µg
|
204
µg
|
| Nylon
cyclone, 1.7 L/min |
85
|
121
|
251
|
100
|
| GK2.69
cyclone, 4.2 L/min |
34
|
49
|
102
|
50
|
| *Eight-hour
sampled masses are combined with results of NIOSH, BOM [1983]. |
|
 |
 |
 |
 |
| |
Table
11
|
| |
|
Table
11. IR method: concentration ranges bracketing applicable exposure
limits for which the NIOSH accuracy criterion is met*
( µg/m3)
|
|
Cyclone
and sampling rate
|
Filter
loading
|
Applicable
exposure limit
|
|
6.72
µg
|
99.7µg
|
161
µg
|
| Nylon
cyclone, 1.7 L/min |
83
|
123
|
198
|
100
|
| GK2.69
cyclone, 4.2 L/min |
34
|
50
|
80
|
50
|
| *Eight-hour
sampled masses are combined with results of NIOSH, BOM [1983]. |
|
 |
 |
 |
 |
| |
|
 |
 |
|
|