Mining Publication: Analysis of Particulate Contamination in Personal Dust Monitor Sampling
Original creation date: June 2008
Authors: JC Volkwein
NIOSHTIC2 Number: 20034163
Proceedings of the 12th U.S./North American Mine Ventilation Symposium, Reno, Nevada, June 9-11, 2008. Wallace, Jr. KG. ed., Reno, NV: University of Nevada - Reno, 2008; :353-358
The National Institute for Occupational Safety and Health (NIOSH) has continued study of the Personal Dust Monitor (PDM) developed by Rupprecht & Patashnick Co. Inc. through a CDC contract and currently being commercialized by Thermo Fisher Scientific. This report examines the effect of oversize particles, cyclone inversion, and cleaning procedures. Clean and exposed filters were examined using light microscopy and particulates counted before and after various instrument manipulations. The data presented show that the dry cleaning procedures do not cause significant (greater than limit of detection) sample contamination by respirable or oversize particulates and that the cyclone is resistant to particulate carry over during inversion.