| DRAFT FOR DISCUSSION
Part 1: Concept for CBRN Air-Purifying Escape Respirator Standard (1) Goal:
Develop a NIOSH standard for escape only air-purifying respirators that
addresses CBRN materials identified as inhalation hazards from possible
terrorist events for use by the general working population. (2) Hazard Categories:
Defining appropriate hazard levels for escape from a possible chemical,
biological, radiological, and nuclear (CBRN) terrorist event is a complex
problem. Analysis of possible escape scenarios indicates the range of
possible hazard concentrations at and between levels typically identified
by emergency responders as the Hot Zone and the Warm Zone. The Hot Zone
is ground zero and can be characterized as the hazard levels associated
with a likely terrorist event, “Most Credible Event” (MCE).
MCE’s for chemical warfare agents (CWA’s) and toxic industrial
materials (TIM’s) expected at a terrorist event are determined using
the Automated Decision Aid System for Hazardous Incidents (ADASHI) modeling
program developed by the U.S. Army Soldier and Biological Chemical Command.
This model considers several parameters associated with the potential
event. These parameters include the means used to transport the CWA or
TIM to the scene, the method of dissemination of the hazard, properties
of the hazard, the quantity of the CWA or TIM used, the availability of
the CWA or TIM, and physical characteristics of the area such as room
size and the degree of ventilation present. Using this approach, challenge
concentrations for sarin gas, GB, and sulfur mustard, HD were determined
to be 2000 mg/m3 for GB and 300 mg/m3 for HD. Similar modeling techniques
are currently being employed for TIM’s that have also been identified
as high threat possibilities.
Warm Zone analysis of the CWA’s and TIM’s are determined
by the immediately dangerous to life or health, IDLH, concentrations or
equivalent for the identified hazards. For GB and HD, the equivalent warm
zone concentrations can be set at 0.19 mg/m3 GB and 2.7 mg/m3 HD, based
on Acute Exposure Guideline Levels (AEGL’s), AEGL 2 values at 30
minutes. Also, high concentrations of some TICs will cause displacement
of oxygen in the contaminated area, thus resulting in an IDLH condition
where the oxygen content falls below 19.5%.
Based on the Hot Zone / Warm Zone GB and HD concentrations, it can be
expected that respirator performance requirements for escape from the
Hot Zone are different from those requirements for escape from or near
Warm Zone concentrations. In addition, the characteristics of the diverse
hazards and buildings or site characteristics vary significantly. No two
are expected to be identical. Because of this, a wide range of strategies
is expected. Certain conditions may involve a dual response strategy:
use of an escape respirator and/or shelter in place. Escape only air-purifying
respirators designed for specific hazards at levels between the Hot and
Warm Zones may be appropriate for specific escape scenarios but do not
represent a universal escape respirator solution for protecting all or
the majority of workers. Furthermore, requirements for acceptable escape
respirator performance for a skyscraper are most likely different than
acceptable escape respirator performance from a 3-level building. The
threat for a metropolitan area located near a major industrial complex,
a chemical plant or oil refinery is not the same as the threat for metropolitan
areas removed from industry.
The concept for escape respirator performance requirements to address
the wide range of variables is segmented into three categories: HIGH,
SPECIFIC, and LOW (General). The categories are associated with a level
of protection as follows:
HIGH: Self-Contained Escape Respirator for unknown conditions
and oxygen deficiency.
SPECIFIC: Air Purifying Escape Respirator for low concentrations
of CWA’s and TIM’s (CBRN Protection) plus high concentrations
of specific TIM’s .
LOW (General): Air Purifying Escape Respirator for low
concentrations of CWA’s and TIM’s (CBRN Protection).
Part 1 of this concept paper addresses the SPECIFIC and LOW (General)
categories for air purifying escape respirators. The HIGH category, self-contained
escape respirator, is addressed in Part 2 of the concept paper.
2(a) Category
vs. Hazard vs. Escape Respirator Type:
Table
1. Escape Respirator Categories
| Category |
Hazard
Description |
Respirator
Type |
HIGH
(Hot & Warm Zones)
|
CWA
& TIM Hazard Threats at High Concentrations and/or Oxygen Deficiency
|
Self
Contained Escape Respirator |
SPECIFIC
(Hot & Warm Zones) |
CWA
+ Specific TIM Hazard Threats at High Concentrations
|
Specific
Gas/Vapor + CWA Air Purifying Escape Respirator |
LOW
(General)
(Warm Zone) |
CWA
& Multiple Hazard Threats at Low Concentrations |
Multi
Gas/Vapor/Particulate Air Purifying Escape Respirator |
2(b) Escape Respirator Multi Gas/Vapor/Particulate
Requirements LOW (General) Category:
Multi Gas/Vapor/Particulate Escape respirators for use at low hazard
threat conditions shall meet the gas/vapor test challenge concentrations
as follows:
| |
Test Concentration (ppm)
Draft |
Breakthrough Concentration (ppm)
Draft |
| Ammonia |
1250 |
150 |
| Cyanogen Chloride |
300 |
0.4 |
| Cyclohexane |
2600 |
10 |
| Formaldehyde |
250 |
10 |
| Hydrogen Cyanide |
470 |
10 |
Hydrogen Sulfide |
500 |
30 |
Nitrogen Dioxide |
100 |
1 ppm NO2 |
Phosgene |
125 |
0.2 |
Phosphine |
150 |
0.5 |
Sulfur Dioxide |
750 |
3 |
(1) Sum of HCN and C2N2
2(c) Escape Respirator Multi Gas/Vapor/Particulate
LOW (General) Category with Carbon Monoxide Requirements:
Escape respirators intended for use at low hazard threat conditions
with carbon monoxide protection shall meet the requirements of paragraph
2(b) plus carbon monoxide as follows: Test Concentration – 3600
ppm; Breakthrough Concentration – 350 ppm.
2(d) Escape Respirator Specific Gas/Vapor/Particulate
Plus CWA Requirements SPECIFIC Category:
Escape respirators intended for use at the specific hazard threat category
conditions shall meet the gas/vapor/particulate testing at identified
conditions of paragraph 2(b) Escape Respirator Multi Gas/Vapor/Particulate
Requirements LOW (General) Category .
Additional specific test agent protections can be added to the minimum
as specified by the applicant for: Ammonia, Cyclohexane, Cyanogen Chloride,
Formaldehyde, Hydrogen Cyanide, Nitrogen Dioxide, Hydrogen Cyanide,
Sulfur Dioxide, Phosgene, Phosphine, and Carbon Monoxide.
2(d) 1. Test
Concentrations for Specific Category:
In addition to the test requirements of paragraph 2(b) Escape Respirator
Multi Gas/Vapor/Particulate Requirements LOW (General) Category test
concentrations for additional specific test agent protections shall
be as specified by the manufacturer.
2(d)2 Breakthrough Concentration for Specific Category:
Test breakthrough concentrations for the specific category shall be
Breakthrough concentrations identified in Section 2(b) Escape Respirator
Multi Gas/Vapor/Particulate Requirements LOW (General) Category .
(3) Respirator Use:
3(a) Escape
Only: Escape respirators are intended to be one time use for escape
only from terrorist events.
3(b) Panic Demand: Each escape respirator shall provide a minimum
duration of 5 minutes when tested at a flow rate of 100 ±10 liters
per minute, 50 ±5 percent relative humidity and 25 ±5°C
for each of the gases/vapors identified in Section 2.
3(c) Duration Rating: Escape respirators will be rated for 15, 30,
45 or 60 minute duration as specified by the manufacturer.
(4) Gas Life Test
Requirements:
4(a) Test Duration:
Test duration will be 15, 30, 45 or 60 minutes as specified by the applicant.
4(b) Gas Life: Gas life tests are performed at room temperature,
25 ±5°C; 25 ±5 percent relative humidity, and 80 ±5
percent relative humidity. Three filters will be tested at each specified
humidity with a flow rate of 64 liters per minute, continuous flow.
Tests will be conducted to the minimum specified service time. Gas testing
shall be performed following environmental conditioning and rough handling.
Service Life testing is performed to the minimum specified service time.
The breakthrough concentration must be no greater than the specified
breakthrough for each tested gas. Testing is terminated after the minimum
specified service time is achieved.
4(c) Particulate Filtration: The filter shall meet the requirements
of a P100 particulate filter as described in 42 CFR, Part 84 paragraphs
84.170, 84.179 and 84.181.
(5) Environmental Conditioning Requirements:
Environmental conditioning
will be performed on escape respirators in the ready-to use configuration.
The ready-to-use configuration is the operational packaging state prior
to use such that immediately upon opening allows the user to don the respirator.
Environmental conditioning shall be performed in accordance with the
following Table:
Durability Test Matrix: Environmental, Transportation and Drop Tests
| Test |
Test Method |
Test Condition |
Duration |
| Hot Constant |
Mil-Std-810F, 501.4 |
71°C
(160°F),
Constant |
5 Weeks |
| Cold Constant |
Mil-Std-810F, 502.4 |
Basic Cold, -32 °C
(-24°F),
Constant |
3 Days |
| Humidity |
Mil-Std-810E, 507.3 |
Realistic, Natural Cycle Humidity Profiles
in the U.S. |
5 Days “quick look”
Mil-Std-810E
Table 507.3-II |
Transportation
Vibration |
MIL-STD-810F, 514.5 |
U. S. Roadway Vibration, Unrestrained |
12 hours/axis, 3 Axes; Total Duration = 36 hours =
12,000 miles |
Drop |
Adopted from NIOSH, CBRN APR Standard |
Height of 3 Feet |
1 Drop on each of the 3 Axes per Unit |
(6) Performance Requirements:
Escape respirator performance requirements considered will include the
following:
6(a) Chemical Agent Permeation
and Penetration Resistance Against Distilled Mustard (HD) and Sarin (GB)
Agent Requirement, LOW (General) and Specific
Category:
The escape respirator system shall resist the permeation and penetration
of distilled sulfur mustard (HD) and Sarin (GB) chemical agents when
tested on an upper-torso manikin connected to a breathing machine operating
at an air flow rate of 40 liters per minute (L/min), 36 respirations
per minute, 1.1 liters tidal volume.
Test requirements for distilled sulfur mustard
(HD) are shown in the following Table:
Table: Simultaneous Liquid and Vapor Challenge of Escape Respirator
with Distilled Sulfur Mustard (HD)
| Agent |
ChallengeConcentration |
Duration of Challenge (min) |
Breathing Machine Airflow Rate (L/min) |
Maximum Peak Excursion (mg/m3) |
Maximum Breakthrough Concentration integrated
over Minimum Service Life (mg-min/m3) |
Number of Systems Tested |
Minimum Service Life (minutes) |
| HD - Vapor |
50 mg/m3 |
15/30/45/60 |
|
|
|
|
|
| HD - Liquid |
0.86 ml |
60 |
40 |
0.60 |
6.0 |
3 |
30/60/90/120 |
(1) Vapor challenge concentration will start immediately
after the liquid drops have been applied and the test chamber has been
sealed.
(2) Liquid volume applied as 25 drops of equal size..
(3) Three consecutive sequential test data points at or exceeding 0.6
mg/m3 will collectively constitute a failure where each test value is
based on a detector sample time of approximately 2 minutes.
(4) The cumulative Ct including all peak data points must not be exceeded
for the duration of the test.
(5) 15, 30, 45 or 60 minutes, equal to the tested duration.
(6) 30, 60, 90 or 120 minutes, twice the tested duration.
Test requirements for Sarin (GB) agent are shown in
the following Table :
Table: Vapor Challenge of Escape Respirator with Sarin (GB).
| Challenge Agent |
Vapor Concentration (mg/m3) |
Vapor Challenge Time (minutes) |
Breathing Machine Airflow Rate (L/min) |
Maximum Peak Excursion (mg/m3) |
Maximum Breakthrough Concentration integrated
over Minimum Service Life (mg-min/m3) |
Number of Systems Tested |
Minimum Service Life (minutes) |
| GB |
210 |
15/30/45/60 |
40 |
0.087 |
2.1 |
3 |
30/60/90/120 |
(1) The vapor challenge concentration generation will
be initiated immediately after test chamber has been sealed.
(2) The test period begins upon initial generation of vapor concentration.
(3) Three consecutive sequential test data points at or exceeding 0.087
mg/m3 will collectively constitute a failure where each test value is
based on a detector sample time of approximately 2 minutes.
(4) The cumulative Ct including all peak data points must not be exceeded
for the duration of the test.
(5) 15, 30, 45 or 60 minutes, equal to the tested duration.
(6) 30, 60, 90 or 120 minutes, twice the tested duration.
6(c) Breathing Resistance:
The resistance of airflow shall be measured at the breathing zone (nosecup
or mouthpiece) of a hood mounted on a head form test apparatus operated
at a continuous airflow rate of 85 liters per minute. The inhalation
resistance shall not exceed 70 mm H2O and the exhalation
resistance shall not exceed 20 mm H2O.
6(d) Breathing Gas::
The carbon dioxide content shall not exceed 2.5% by volume and the oxygen
concentration shall not be less than 19.5% by volume when tested on
an Automated Breathing Metabolic Simulator (ABMS) operated per the requirements
of the following table:
Metabolic Variables for the ABMS Breathing Gas
Test
Work
Rate
|
f (breath/min) |
Vt (Liters, STPD) |
VE
(L/min,
STPD) |
VO2
(L/min,
STPD) |
VCO2
(L/min,
STPD) |
R |
ABMS
Waveform |
Approx.
Peak
Flows
(L/min,
BTPS |
| 1 |
12 |
0.8 |
10.0 |
0.5 |
0.4 |
0.80 |
BS1210 |
45.0 |
2 |
19.5 |
1.30 |
25.3 |
1.0 |
0.8 |
0.82 |
BS1925 |
96.6 |
3 |
28.0 |
1.35 |
38.0 |
1.5 |
1.3 |
0.87 |
BS2838 |
149.5 |
4 |
32.6 |
1.90 |
62.0 |
2.0 |
1.9 |
0.95 |
BS3362 |
243.8 |
| 5 |
34.2
|
2.00 |
70.0 |
2.5 |
2.5 |
1.00 |
BD3470 |
209.3ex
241.5in |
6 |
36.4 |
2.20 |
80.0 |
3.0 |
3.15 |
1.05 |
BD3680 |
246.1ex
276.0in |
Test time at each work rate shall be 10 minutes.
6(e) Communications (Speech Intelligibility):
The Communication (Speech Intelligibility) capability is an optional
feature for the AP Escape Respirator. The manufacturer will inform NIOSH
whether they want their respirator certified with the Communication
(Speech Intelligibility) Endorsement. If the manufacturer declines to
have the Communication Endorsement, then the manufacturer’s AP
Escape Respirator will not be required to meet the Communication Performance
Requirement. If the manufacturer informs NIOSH that they want the Communication
Endorsement, then the respirator must meet or exceed the Communication
Performance Requirement.
Communication Performance Requirement is based upon performance using
a Modified Rhyme Test (MRT). The Communications Performance Requirement
is met if the overall performance rating is greater than or equal to
seventy (70) percent when tested in accordance with NIOSH Standard Test
Procedure CET-APRS-STP-CBRN-0313. The MRT will be performed with a steady
background noise of 60 dBA consisting of a broadband “pink”
noise. The distance between the listeners and speakers shall be 10 feet
(3 m).
6(f) Field of View:
The CBRN AP Escape Respirator shall obtain a Visual Field Score (VFS)
of 70 or greater when tested in accordance with NIOSH Standard Test
Procedure CET-APRS-STP-CBRN-0314. The VFS shall be obtained by using
a medium size respirator or equivalent that is sized to fit the Head
Form described in Figure 14 of EN 136, Respiratory protective devices
– Full face masks – Requirements, testing, marking; January
1998 or equivalent.
The VFS is determined by using a VFS grid (Dots on visual field) as
defined in the American Medical Association Guides to the Evaluation
of Permanent Impairment, 5th Edition (2000) that is overlaid on
the diagram of the visual field plot obtained using the spherical shell
of EN 136 apertometer or equivalent. The VFS score is the average of
three fittings of the same respirator on the specified head form.
From Table-7, Ranges of Field Loss and O&M (Orientation and Mobility)
Ability of the Visual Standards Report, Aspects and Ranges of Vision
Loss with Emphasis on Population Surveys, International Council of Ophthalmology
at the 29th International Congress of Ophthalmology Sydney Australia,
April 2002, the VFS score of 70 points is the minimum value which falls
into the Mild Visual Impairment category. This category translates to
a functional ability of Normal O&M Performance that needs more scanning,
occasionally surprised by events on the side. If the visual field is
mildly impaired, the user has scanning capability.
6(g) Donning:
The time to fully don the respirator from the ready-to use configuration
shall be no greater than 30 seconds. The ready-to-use configuration
is the operational packaging state prior to use such that immediately
upon opening allows the user to don the respirator.
6(h) Fogging:
The AP Escape Respirator shall demonstrate an average Visual Acuity
Score (VAS) of greater or equal to 70 points for all measurements for
each individual. The wearer shall not experience undue discomfort because
of restrictions to breathing or other physical or chemical changes to
the respirator.
The respirator shall be donned by the test subject in an indoor ambient
temperature of approximately 72°F +/- 2° F at 30% RH +/- 5%
and then shall enter into a simulated outdoor extreme temperature chamber
where the visual acuity tests shall be administered. The APR Escape
Respirator shall be tested for fogging in the hot/humid condition of
90 °F/ 60% RH and the cold condition of 13.1°F.
From Table-5, Ranges of Reading Ability, of the Visual Standards Report,
Aspects and Ranges of Vision Loss with Emphasis on Population Surveys,
International Council of Ophthalmology at the 29th International Congress
of Ophthalmology Sydney Australia, April 2002, the VAS score of 70 points
indicates Statistical estimates of Reading Ability that borderlines
between the Normal Reading Speed/Reduced Reading Distance and the Near-Normal
with Appropriate Reading Aids Category.
6(i) Flammability and Heat Resistance:
LOW (General) and SPECIFIC Category Escape Respirator shall be tested
for Flammability and Heat Resistance using the test equipment specified
in EN 136, Respiratory Protective Devices, Full Face Masks, Requirements,
testing, Marking, 1998 Edition. No component of the respirator shall
have an afterflame after 5 seconds. No component of the escape respirator
shall drip, melt, or develop a visible hole or damaged in any manner
that compromises the breathing protection provided by the respirator.
The distance between the outer surface of the escape respirator and
the burner tips shall be adjusted to 250 mm ± 6.4 mm. The pressure
reducer shall be adjusted to 2.1 bar ± .05 psi. The temperature
of the flame positioned 250 mm ± 6.4 mm above the burner tip
shall be 800 °C ± 50 ° C. The respirator shall be rotated
once through the flame at a velocity of 6 ± 0.5 cm/s. Where components
of the respirator such as valves, filters, etc. are arranged on the
respirator, the test shall be repeated with these components at the
appropriate height of 250 mm ± 6.4 mm.
6(k) Laboratory Respirator Protection Level:
The measured laboratory respirator protection level (LRPL) for each
air purifying escape respirator shall be 2000, when the respirator is
tested in a negative pressure mode in an atmosphere containing 20-40
mg/m3 corn oil aerosol of a mass median aerodynamic diameter
of 0.4 to 0.6 micrometers.
| |
Small |
Medium |
Large |
| Head Circumference |
NA |
NA |
576 - 600 |
| Neck Circumference |
307 - 350 |
351 - 375 |
376 - 409 |
| Face Length |
NA |
NA |
124 – 133.5 |
Five respirators shall be tested with test subjects from each cell
of the above table with dimensions identified. Each cell is uniquely
tested.
(7) Design Considerations:
7(a) General:
The escape respirator shall provide a barrier from ambient conditions
for the wearer’s entire head, eyes, and respiratory system. The
escape respirator shall not require the use of hands to maintain the
respirator position to ensure proper function of the respirator when
fully donned.
7(b) Hood Type Device:
The escape respirator shall be designed as a hooded device. The hood
shall include an area for field vision and shall be compatible with
wearing of glasses.
7(c) Respiratory Protection System:
The respiratory protection system shall consist of an oral/nasal cup
or mouthpiece. If a mouthpiece is employed, a method of preventing nasal
breathing must be provided. The respiratory protection system shall
be designed such that the air purifying filter cannot be degraded by
the carbon dioxide and humidity of the exhaled gas.
(8) 42 CFR Applicable Sections:
The following sections of 42 CFR, Part 84 are applicable:
42 CFR, Part 84, Subparts A, B, D, E, F, and G:
Subpart A: General Provisions
Subpart B: Application for Approval
Subpart D: Approval and Disapproval
Subpart E: Quality Control
Subpart F: Classification of Approved Respirators
Subpart G: General Construction and Performance
42 CFR, Part 84, Subpart K; the following paragraphs apply:
84.170 Non-powered air purifying particulate respirators; description
84.179 Non-powered air purifying particulate respirators; filter identification
84.180 Non-powered air purifying particulate filter efficiency
(9) Service and Maintenance:
TBD
(10) Training:
TBD
(11) Cautions and Limitations:
TBD
Part 2: Concept for CBRN Self-Contained Escape Respirator Standard
(1) General:
The concept for escape respirator performance requirements to address
the wide range of variables is segmented into three categories: HIGH,
SPECIFIC, and LOW (General). The categories are associated with a level
of protection as follows:
HIGH: Self-Contained Escape Respirator for unknown conditions
and oxygen deficiency.
SPECIFIC: Air Purifying Escape Respirator for high concentrations
of CWA’s and specific TIM’s.
LOW (General): Air Purifying Escape Respirator for low
concentrations of CWA’s and TIM’s.
The standard discussed in Part 2 of this concept paper addresses the
HIGH category for self-contained escape respirators. The SPECIFIC and
LOW (General) Categories are discussed in Part 1 of the Escape Respirator
Concept.
(2) Requirements for the HIGH Category Self-Contained Escape
respirator are identified as a three tier set of requirements:
- 42 CFR, Part 84, Subpart H Escape Respirator Approval
- Enhanced Escape Respirator Performance Requirements
- CBRN Requirements.
(3) 42 CFR, Part 84, Subpart H Approval:
The HIGH Category Escape Respirator must be NIOSH approved as a self-contained
escape respirator for a service life of 15, 30, 45 or 60 minutes duration
in accordance with the requirements of 42 CFR, Part 84, Subpart H.
(4) Enhanced Escape Respirator Performance Requirements:
Escape respirators will be environmentally conditioned prior to conducting
test and evaluations specified in sections 4 and 5.
(4.1) Environmental Conditioning:
Environmental conditioning will be performed on escape respirators
in the ready-to-use configuration. The ready-to-use configuration is
the operational packaging state prior to use such that immediately upon
opening allows the user to don the respirator.
Environmental conditioning shall be performed in accordance with the
following Table:
Durability Test Matrix: Environmental, Transportation and Drop
Tests
Test |
Test Method |
Test Condition |
Duration |
| Hot Constant |
MIL-STD-810F, 501.4 , |
71 0C (160 0F) |
Constant 5 Weeks |
| Cold Constant |
MIL-STD-810F, 502.4 |
Basic Cold, -32 0C (-24 0F), |
Constant 3 Days |
| Humidity |
MIL-STD-810E, 507.3 |
Realistic, Natural Cycle Humidity Profiles in the U.S. |
5 Days “quick look”
Mil-Std-810E
Table 507.3-II |
Transportation
Vibration |
MIL-STD-810F, 514.5 |
U. S. Roadway Vibration, Unrestrained |
12 hours/axis, 3 Axes; Total Duration = 36 hours = 12,000 miles |
Drop |
Adopted from NIOSH, CBRN APR Standard |
Height of 3 Feet |
1 Drop on each of the 3 Axes per Unit |
(4.2) Hood:
The escape respirator shall be designed as a hooded device. The hood
shall include an area for field of vision and shall be compatible with
wearing glasses. The escape respirator shall not require the use of
hands to maintain the respirator position to ensure proper function
of the respirator when fully donned.
(4.3) Respiratory Protection System:
The respiratory protection system shall consist of an oral/nasal cup
or mouthpiece. If a mouthpiece is employed a method of preventing nasal
breathing must be provided.
(4.4) Dermal Protection:
The HIGH Category Escape Respirator shall include a means of full body
protection in the form of a cape, apron or coat that may be packaged
separate from the respirator. The cape, apron or coat shall cover the
shoulders, arms, upper body and legs.
(4.5) Donning Time:
The HIGH Category Escape Respirator shall be fully donned and activated
from its stored configuration in less than 30 seconds.
(4.6) Flammability and Heat Resistance:
HIGH Category Escape Respirator shall be tested for Flammability and
Heat Resistance using the test equipment specified in EN 136, Respiratory
Protective Devices, Full Face Masks, Requirements, testing, Marking,
1998 Edition. No component of the respirator shall have an afterflame
after 5 seconds. No component of the escape respirator shall drip, melt,
or develop a visible hole or damaged in any manner that compromises
the breathing protection provided by the respirator.
The distance between the outer surface of the escape respirator and
the burner tips shall be adjusted to 250 mm ± 6.4 mm. The pressure
reducer shall be adjusted to 2.1 bar ± .05 psi. The temperature
of the flame positioned 250 mm ± 6.4 mm above the burner tip
shall be 80° 0 C ± 50°C. The respirator shall be rotated
once through the flame at a velocity of 6 ± 0.5 cm/s. Where components
of the respirator such as valves, filters, etc. are arranged on the
respirator, the test shall be repeated with these components at the
appropriate height of 250 mm ± 6.4 mm. If compressed oxygen is
used in the escape respirator this requirement will be tested using
a surrogate oxygen pressure vessel.
(4.7) Field of View:
The CBRN Self Contained Escape Respirator shall obtain a Visual Field
Score (VFS) of 70 or greater when tested in accordance with NIOSH Standard
Test Procedure CET-APRS-STP-CBRN-0314. The VFS shall be obtained by
using a medium size respirator or equivalent that is sized to fit the
Head Form described in Figure 14 of EN 136, Respiratory protective devices
– Full face masks – Requirements, testing, marking; January
1998 or equivalent.
The VFS is determined by using a VFS grid (Dots on visual field) as
defined in the American Medical Association Guides to the Evaluation
of Permanent Impairment, 5th Edition (2000) that is overlaid on the
diagram of the visual field plot obtained using the spherical shell
of EN 136 apertometer or equivalent. The VFS score is the average of
three fittings of the same respirator on the specified head form.
From Table-7, Ranges of Field Loss and O&M (Orientation and Mobility)
Ability of the Visual Standards Report, Aspects and Ranges of Vision
Loss with Emphasis on Population Surveys, International Council of Ophthalmology
at the 29th International Congress of Ophthalmology Sydney Australia,
April 2002, the VFS score of 70 points is the minimum value which falls
into the Mild Visual Impairment category. This category translates to
a functional ability of Normal O&M Performance that needs more scanning,
occasionally surprised by events on the side. If the visual field is
mildly impaired, the user has scanning capability.
(4.8) Fogging:
The CBRN Self Contained Escape Respirator shall demonstrate an average
Visual Acuity Score (VAS) of greater or equal to 70 points for all measurements
for each individual. The wearer shall not experience undue discomfort
because of restrictions to breathing or other physical or chemical changes
to the respirator.
The respirator shall be donned by the test subject in an indoor ambient
temperature of approximately 72°F +/- 2° F at 30% RH +/- 5%
and then shall enter into a simulated outdoor extreme temperature chamber
where the visual acuity tests shall be administered. The APR Escape
Respirator shall be tested for fogging in the hot/humid condition of
90 0F/ 60% RH and the low temperature use limit for which the respirator
is approved as part of 42 CFR, Part 84, Subpart H.
From Table-5, Ranges of Reading Ability, of the Visual Standards Report,
Aspects and Ranges of Vision Loss with Emphasis on Population Surveys,
International Council of Ophthalmology at the 29th International Congress
of Ophthalmology Sydney Australia, April 2002, the VAS score of 70 points
indicates Statistical estimates of Reading Ability that borderlines
between the Normal Reading Speed/Reduced Reading Distance and the Near-Normal
with Appropriate Reading Aids Category.
(4.9) Breathing Gas Concentrations:
The inhaled CO2 concentration shall be less than 1.5% by volume and
the O2 concentration greater than 19.5% by volume when tested with human
test subjects in accordance with Man Test 1, 42 CFR, Subpart H, Paragraph
84.100 work activities. The concentration of CO2 and O2 in the inspired
gas at the mouth will be measured continuously throughout the test.
Human subject tests will be performed with two test subjects at each
work rate, one weighing 60 kilograms or less and one weighing 80 kilograms
or more. Tests shall be performed for the rated duration of the respirator
as determined by 3(c) Duration Rating for each work rate specified.
NIOSH will also perform machine tests with an Automated Breathing Metabolic
Simulator (ABMS) prior to performing human subject testing to assess
the function of the respirator to determine it’s suitability for
human subject testing. The ABMS tests will be performed at the following
conditions:
Work
Rate
|
f (breath/min) |
Vt (Liters, STPD) |
VE
(L/min,
STPD) |
VO2
(L/min,
STPD) |
VCO2
(L/min,
STPD) |
R |
ABMS
Waveform |
Approx.
Peak
Flows
(L/min,
BTPS |
| 1 |
12 |
0.8 |
10.0 |
0.5 |
0.4 |
0.80 |
BS1210 |
45.0 |
2 |
19.5 |
1.30 |
25.3 |
1.0 |
0.8 |
0.82 |
BS1925 |
96.6 |
3 |
28.0 |
1.35 |
38.0 |
1.5 |
1.3 |
0.87 |
BS2838 |
149.5 |
4 |
32.6 |
1.90 |
62.0 |
2.0 |
1.9 |
0.95 |
BS3362 |
243.8 |
| 5 |
34.2
|
2.00 |
70.0 |
2.5 |
2.5 |
1.00 |
BD3470 |
209.3ex
241.5in |
6 |
36.4 |
2.20 |
80.0 |
3.0 |
3.15 |
1.05 |
BD3680 |
246.1ex
276.0in |
Test time at each work rate shall be the rated duration of the respirator.
(5.0) CBRN Requirements:
(5.1) Laboratory Respiratory Protection Level (LRPL):
The measured laboratory respirator protection level (LRPL) for each
High Category Escape Respirator air purifying escape respirator shall
be 2000, when the respirator is tested in a negative pressure mode in
an atmosphere containing 20-40 mg/m3 corn oil aerosol of
a mass median aerodynamic diameter of 0.4 to 0.6 micrometers.
| |
Small |
Medium |
Large |
| Head Circumference |
NA |
NA |
576 - 600 |
| Neck Circumference |
307 - 350 |
351 - 375 |
376 - 409 |
| Face Length |
NA |
NA |
124 – 133.5 |
Five respirators shall be tested with test subjects from each cell
of the above table with representative dimensions. Each cell is uniquely
tested.
5.2) Live Agent Test:
(1). Chemical Agent Permeation and Penetration Resistance
Against Distilled Mustard (HD) and Sarin (GB) Agent Test Requirement
Open-circuit, positive-pressure SCBAs, including all components and
accessories except the air cylinder (shell), shall resist the permeation
and penetration of distilled sulfur mustard (HD) and sarin (GB) chemical
agents when tested on an upper-torso manikin connected to a breathing
machine operating at an air flow rate of 40 liters per minute (L/min),
36 respirations per minute, 1.1 liters tidal volume.
Test requirements for distilled sulfur mustard (HD) are shown in Table
1.
Table 1: Simultaneous Liquid and Vapor Challenge of SCBA with
Distilled Sulfur Mustard (HD)
| Agent |
ChallengeConcentration |
Duration of Challenge (min) |
Breathing Machine Airflow Rate (L/min) |
Maximum Peak Excursion (mg/m3) |
Maximum Breakthrough (concentration integrated
over Minimum Service Life) (mg-min/m3) |
Number of Systems Tested |
Minimum Service Life (minutes) |
| HD - Vapor |
300 mg/m3 |
Stated Duration |
40 |
0.60 |
6.0 |
3 |
Twice Stated Duration |
| HD - Liquid |
0.86 ml |
360 |
(1) Vapor challenge concentration will start immediately after the
liquid drops have been applied and the test chamber has been sealed.
(2) The test period begins upon start of initial vapor generation.
(3) Three consecutive sequential test data points at or exceeding 0.6
mg/m3 will collectively constitute a failure where each test value is
based on a detector sample time of approximately 2 minutes.
(4) The cumulative Ct including all peak data points must not be exceeded
for the duration of the 6-hour test.
(5) For a duration period equal to the stated duration life.
(6) For a duration period equal to twice the stated service life.
Test requirements for sarin (GB) agent are shown in Table 2.
Table 2: Vapor Challenge of SCBA with Sarin (GB)
| Challenge Agent |
Vapor Concentration (mg/m3) |
Vapor Challenge Time (minutes) |
Breathing Machine Airflow Rate (L/min) |
Maximum Peak Excursion (mg/m3) |
Maximum Breakthrough (concentration integrated
over Minimum Service Life) (mg-min/m3) |
Number of Systems Tested |
Minimum Service Life (minutes) |
| GB |
2,000 mg/m3 |
Stated Duration |
40 |
0.087 |
2.1 |
3 |
Twice Stated Duration |
(1) The vapor challenge concentration generation will be initiated
immediately after test chamber has been sealed.
(2) The test period begins upon initial generation of vapor concentration.
(3) Three consecutive sequential test data points at or exceeding 0.087
mg/m3 will collectively constitute a failure where each test value is
based on a detector sample time of approximately 2 minutes.
(4) The cumulative Ct including all peak data points must not be exceeded
for the duration of the 6-hour test.
(5) For a duration period equal to the stated duration life.
(6) For a duration period equal to twice the stated service life.
|