Infection Prevention during Blood Glucose Monitoring and Insulin Administration

Summary

The Centers for Disease Control and Prevention (CDC) has become increasingly concerned about the risks for transmitting hepatitis B virus (HBV) and other infectious diseases during assisted blood glucose (blood sugar) monitoring and insulin administration.

CDC is alerting all persons who assist others with blood glucose monitoring and/or insulin administration of the following infection control requirements:

  • Fingerstick devices should never be used for more than one person
  • Whenever possible, blood glucose meters should not be shared. If they must be shared, the device should be cleaned and disinfected after every use, per manufacturer’s instructions. If the manufacturer does not specify how the device should be cleaned and disinfected then it should not be shared.
  • Insulin pens and other medication cartridges and syringes are for single-patient-use only and should never be used for more than one person
Blood Glucose Monitoring and Insulin Administration
Do not share your insulin pen.

Monitoring of blood glucose levels is frequently performed to guide therapy for persons with diabetes. Blood glucose monitoring and insulin administration can be accomplished in two ways: self-monitoring of blood glucose and insulin administration, where the individual performs all steps of the testing and insulin administration themselves, and assisted monitoring of blood glucose and insulin administration, where another person assists with or performs testing and insulin administration for an individual.

Examples of settings where assisted monitoring of blood glucose and insulin administration may occur include:

  • Hospitals or clinics
  • Long term care settings such as nursing homes and assisted living facilities
  • Senior centers
  • Health fairs
  • Correctional facilities
  • Schools or camps
Unsafe Practices during Blood Glucose Monitoring and Insulin Administration

An underappreciated risk of blood glucose testing is the opportunity for exposure to bloodborne viruses (HBV, hepatitis C virus, and HIV) through contaminated equipment and supplies if devices used for testing and/or insulin administration (e.g., blood glucose meters, fingerstick devices, insulin pens) are shared.

Outbreaks of hepatitis B virus (HBV) infection associated with blood glucose monitoring have been identified with increasing regularity, particularly in long-term care settings, such as nursing homes and assisted living facilities, where residents often require assistance with monitoring of blood glucose levels and/or insulin administration. In the last 10 years, alone, there have been at least 15 outbreaks of HBV infection associated with providers failing to follow basic principles of infection control when assisting with blood glucose monitoring. Due to under-reporting and under recognition of acute infection, the number of outbreaks due to unsafe diabetes care practices identified to date are likely an underestimate.

Although the majority of these outbreaks have been reported in long-term care settings, the risk of infection is present in any setting where blood glucose monitoring equipment is shared or those assisting with blood glucose monitoring and/or insulin administration fail to follow basic principles of infection control. For example, at a health fair in New Mexico in 2010, dozens of attendees were potentially exposed to bloodborne viruses when fingerstick devices were inappropriately reused for multiple persons to conduct diabetes screening. Additionally, at a hospital in Texas in 2009, more than 2,000 persons were notified and recommended to undergo testing for bloodborne viruses after individual insulin pens were used for multiple persons.

Unsafe practices during assisted monitoring of blood glucose and insulin administration that have contributed to transmission of HBV or have put persons at risk for infection include:

  • Using fingerstick devices for more than one person
  • Using a blood glucose meter for more than one person without cleaning and disinfecting it in between uses
  • Using insulin pens for more than one person
  • Failing to change gloves and perform hand hygiene between fingerstick procedures
Best Practices for Assisted Blood Glucose Monitoring and Insulin Administration

The following are infection control recommendations that anyone who performs or assists with blood glucose monitoring and /or insulin administration should review to assure they are not placing themselves or persons in their care at risk.

These recommendations apply not only to licensed healthcare facilities but also to any setting where fingerstick procedures are performed and/or insulin is administered, including assisted living or residential care facilities, clinics, health fairs, shelters, detention facilities, schools, and camps. Protection from bloodborne viruses and other infections is a basic requirement and expectation anywhere healthcare is provided.

Fingerstick Devices

Fingerstick devices, also called lancing devices, are devices that are used to prick the skin and obtain drops of blood for testing. There are two main types of fingerstick devices: those that are designed for reuse on a single person and those that are disposable and for single-use.

  • Reusable Devices: These devices often resemble a pen and have the means to remove and replace the lancet after each use, allowing the device to be used more than once. Some of these devices have been previously approved and marketed for multi-patient use, and require the lancet and disposable components (platforms or endcaps) to be changed between each patient. However, due to failures to change the disposable components, difficulties with cleaning and disinfection after use, and their link to multiple HBV infection outbreaks, CDC recommends that these devices never be used for more than one person. If these devices are used, it should only be by individual persons using these devices for self-monitoring of blood glucose.
  • Single-use, auto-disabling fingerstick devices: These are devices that are disposable and prevent reuse through an auto-disabling feature. In settings where assisted monitoring of blood glucose is performed, single-use, auto-disabling fingerstick devices should be used.
Blood Glucose Meters

Blood glucose meters are devices that measure blood glucose levels.

  • Whenever possible, blood glucose meters should be assigned to an individual person and not be shared.
  • If blood glucose meters must be shared, the device should be cleaned and disinfected after every use, per manufacturer’s instructions, to prevent carry-over of blood and infectious agents.  If the manufacturer does not specify how the device should be cleaned and disinfected then it should not be shared.
A simple rule for safe care:

 

If shared, blood glucose meters should be cleaned and disinfected after every use.

Recommended Practices for Preventing Bloodborne Pathogen Transmission during Blood Glucose Monitoring and Insulin Administration in Healthcare Settings
  • Fingerstick Devices
    • Restrict use of fingerstick devices to individual persons. They should never be used for more than one person. Select single-use lancets that permanently retract upon puncture. This adds an extra layer of safety for the patient and the provider.
    • Dispose of used lancets at the point of use in an approved sharps container. Never reuse lancets.
  • Blood Glucose Meters
    • Whenever possible, blood glucose meters should be assigned to an individual person and not be shared.
      • If blood glucose meters must be shared, the device should be cleaned and disinfected after every use, per manufacturer’s instructions, to prevent carry-over of blood and infectious agents. If the manufacturer does not specify how the device should be cleaned and disinfected then it should not be shared.
  • General
    • Unused supplies and medications should be maintained in clean areas separate from used supplies and equipment (e.g., glucose meters). Do not carry supplies and medications in pockets.
Insulin Administration
  • Insulin pens should be assigned to individual persons and labeled appropriately. They should never be used for more than one person.
  • Multiple-dose vials of insulin should be dedicated to a single person whenever possible.
    • If the vial must be used for more than one person it should be stored and prepared in a dedicated medication preparation area outside of the patient care environment and away from potentially contaminated equipment
    • Medication vials should always be entered with a new needle and new syringe
    • Dispose of used injection equipment at point of use in an approved sharps container. Never reuse needles or syringes.
Hand Hygiene (Hand washing with soap and water or use of an alcohol-based hand rub)
  • Wear gloves during blood glucose monitoring and during any other procedure that involves potential exposure to blood or body fluids.
  • Change gloves between patient contacts. Change gloves that have touched potentially blood-contaminated objects or fingerstick wounds before touching clean surfaces. Discard gloves in appropriate receptacles.
  • Perform hand hygiene immediately after removal of gloves and before touching other medical supplies intended for use on other persons.
Training and Oversight
  • Review regularly individual schedules for persons requiring assistance with blood glucose monitoring and/or insulin administration.
  • Provide a full hepatitis B vaccination series to all previously unvaccinated staff persons whose activities involve contact with blood or body fluids.
  • Establish responsibility for oversight of infection control activities. Provide staff members who assume responsibilities for fingersticks and injections with infection control training.
  • Assess adherence to infection control recommendations for blood glucose monitoring and insulin administration by periodically observing staff who perform or assist with these procedures and tracking use of supplies.
  • Report to public health authorities any suspected instances of a newly acquired bloodborne infection, such as hepatitis B, in a patient, facility resident, or staff member.
  • Check with state authorities for specific state and federal regulations regarding laboratory testing.
References

For more information regarding bloodborne pathogen transmission associated with unsafe practices during assisted monitoring of blood glucose, consult the following resources.

Centers for Disease Control and Prevention. Nosocomial hepatitis B virus associated with spring-loaded fingerstick blood sampling devices – Ohio and New York City, 1996MMWR 1997;46:217-221.

Centers for Disease Control and Prevention. Notes from the Field: Deaths from Acute Hepatitis B Virus Infection Associated with Assisted Blood Glucose Monitoring in an Assisted-Living Facility – North Carolina, August-October 2010MMWR 2011;60:182.

Centers for Disease Control and Prevention. Transmission of hepatitis B virus among persons undergoing blood glucose monitoring in long-term-care facilities–Mississippi, North Carolina, and Los Angeles County, California, 2003-2004MMWR 2005;54:220-223.

Counard CC, Perz JF, Linchancgo PC, et al. Acute hepatitis B outbreaks related to fingerstick blood glucose monitoring in two assisted living facilities.external icon J Am Geriatric Soc 2010;58:306-311.

Khan AJ, Cotter SM, Schulz B, et al. Nosocomial transmission of hepatitis B virus infection among residents with diabetes in a skilled nursing facility. Infect Control Hosp Epidemiol 2002;23:313-318.

Klonoff DG. Improving the Safety of Blood Glucose Monitoring. J Diabetes Sci Technology 2011;5(6): 1307-1311

Klonoff DC, Perz JF. Assisted Monitoring of Blood Glucose: Special Safety Needs for a New Paradigm in Testing Glucose external iconJ Diabetes Sci Technol2010;4(5):1027-1031

Patel AS, White-Comstock MB, Woolard D, Perz JF. Infection Control Practices in Assisted Living Facilities: A Response to Hepatitis B Virus Infection Outbreaks. Infect Control Hosp Epidemiol 2009;30:209-214.

Polish LB, Shapiro CN, Bauer F, Klotz P, Ginier P, Roberto RR, Margolis HS, Alter MJ. Nosocomial transmission of hepatitis B virus associated with a spring-loaded finger-stick device. N Engl J Med 1992;326(11):721–5.

Quale JM, Landman D, Wallace B, Atwood E, Ditore V, Frucher G. Deja vu: nosocomial hepatitis B virus transmission and fingerstick monitoring. Am J Med1998;105(4):296–301.

Southwick KL, Clement EJ, Konings F, VanZetta S, Johnson GS, Schaffzin JK. Cluster of hepatitis B infection among residents of an assisted living facility–New York, 2007. In Final Program and Abstracts. Sixth International Conference on Emerging Infectious Diseases. Atlanta, GA. March 16-19, 2008.

Thompson ND, Barry V, Alelis K, Gui D, Perz JF. Evaluation of the Potential for Bloodborne Pathogen Transmission Associated with Diabetes Care Practices in Nursing Homes and Assisted Living Facilities, Pinellas County. JAGS 2010;58:914-918.

Thompson ND, Perz JF. Eliminating the Blood: Ongoing Outbreaks of Hepatitis B Virus Infection and the Need for Innovative Glucose Monitoring Techniques. [PDF 445 KB]external iconJ Diabetes Sci Technol 2009;3(2):283-288.

Thompson ND, Schaefer MK. “Never Events”: Hepatitis B Outbreaks and Patient Notifications Resulting from Unsafe Practices during Assisted Monitoring of Blood Glucose, 2009-2010. [PDF 1,118 KB]external icon J Diabetes Sci Technol 2011;5(6):1396-1402.