NIOSHTIC-2 Publications Search

Determine the Reliability of Passive Dosimeters.

Authors
Parker-CD
Source
NIOSH 1976 Jan:44 pages
Link
NIOSHTIC No.
00052626
Abstract
Nitrogen-dioxide (10102440), sulfur-dioxide (7446095) and hydrogen- sulfide (7783064) dosimeters are exposed to controlled environments to determine their accuracies and the effects of temperature, humidity, and face velocity upon their responses to test conditions. The responses of these dosimeters to ozone (10028156) and nitrogen- oxide are also measured. Test apparatus and procedures and the experimental results are described with brief notes on the dosimeters.
Keywords
NIOSH-Contract; Contract-210-75-0080; Toxic-gases; Air-pollution; Air-sampling; Instrumentation; Air-contamination; Air-quality-control; Air-monitoring;
CAS No.
10102-44-0; 7446-09-5; 7783-06-4; 10028-15-6;
Publication Date
19760101
Funding Type
Contract;
Fiscal Year
1976
Identifying No.
Contract-210-75-0080
NIOSH Division
DLCD;
Source Name
Division of Laboratories and Criteria Development, NIOSH, Cincinnati, Ohio