Table 3. Performance Tests to be Applied to the Three Classes of Biological Safety Cabinets

Tests Performed For Biosafety Cabinet
Class
I
Class
II
Class
III
Primary Containment      
Cabinet Integrity N/A A A
HEPA Filter Leak Req Req Req
Downflow Velocity Profile A Req Req
Negative Pressure/Ventilation Rate B A Req
Airflow Smoke Patterns Req Req E/F
Alarms and Interlocks C, D C, D Req
Electrical Safety      
Electrical Leakage, etc. E, D E, D E, D
Ground Fault Interupter D D D
Other    
Lightning Intensity E E E
UV Intensity C, E C, E C, E
Noise Level E E E
Vibration E E E
Req Required during certification.
A Required for proper certification if the cabinet is new, has been moved or panels have been removed for maintenance.
B If used with gloves.
C If present.
D Encouraged for electrical safety.
E Optional, at the discretion of the user.
F Used to determine air distribution within cabinet for clean to dirty procedures.
N/A Not applicable