Table 3. Performance Tests to be Applied to the Three Classes of Biological Safety Cabinets
Tests Performed For Biosafety Cabinet Class
IClass
IIClass
IIIPrimary Containment Cabinet Integrity N/A A A HEPA Filter Leak Req Req Req Downflow Velocity Profile A Req Req Negative Pressure/Ventilation Rate B A Req Airflow Smoke Patterns Req Req E/F Alarms and Interlocks C, D C, D Req Electrical Safety Electrical Leakage, etc. E, D E, D E, D Ground Fault Interupter D D D Other Lightning Intensity E E E UV Intensity C, E C, E C, E Noise Level E E E Vibration E E E
Req Required during certification. A Required for proper certification if the cabinet is new, has been moved or panels have been removed for maintenance. B If used with gloves. C If present. D Encouraged for electrical safety. E Optional, at the discretion of the user. F Used to determine air distribution within cabinet for clean to dirty procedures. N/A Not applicable