Skip directly to search Skip directly to A to Z list Skip directly to page options Skip directly to site content

NIOSHTIC-2 Publications Search

Search Results

An evaluation of a Porapak Q column at cryogenic temperatures for the separation of helium-3, neon, and hydrogen in helium-4.

Authors
Weems-GW; Hoffman-CA; Howard-AR
Source
J Chromatogr Sci 1971 Jul; 9(7):444-446
NIOSHTIC No.
10010634
Abstract
An investigation was undertaken to evaluate a Porapak Q column for separating helium-3, neon, and hydrogen in helium-4. The column temperatures used for this study were 194 and 76 K. The results of this investigation were that helium-3, neon, and hydrogen are separated by using a 4.9 m 0.47 cm, i.d., column filled with 80/100-mesh Porapak Q using a helium-4 carrier gas. Using the separation characteristics of this column material, a chromatographic analytical method was devised.
CODEN
JCHSBZ
Publication Date
19710701
Document Type
OP; Journal Article
Fiscal Year
1971