Skip directly to search Skip directly to A to Z list Skip directly to page options Skip directly to site content

NIOSHTIC-2 Publications Search

Search Results

Characteristics of a Radon Diffusion Chamber With Electrical Collection Using Plastic Nuclear Track Detectors.

Authors
Frank-AL; Benton-EV
Source
NTIS: PB/84-184209 Available for Reference At Bureau of Mines Libraries :17 pages
Link
NIOSHTIC No.
10003527
Abstract
A portable radon diffusion chamber was tested with electrical collection of radon daughter nuclei for enhancement of sensitivity. The detector was small enough to be worn suspended from a belt. With a battery-supplied high voltage, the sensitivity was found to be 1.43 Tracks/cm2 per pci-h/l of ambient radon at 50% relative humidity. The track densities on the surfaces of the cr-39 plastic nuclear track detectors were found to be nonuniform, making the determinations of average densities with high accuracy difficult. The fabrication of electrets for use in the detector was investigated in accordance with published techniques with negative results.
Publication Date
19830101
Document Type
CP; Final Contract Report;
Fiscal Year
1983
NTIS Accession No.
PB84-184209
NTIS Price
A03
Identifying No.
OFR 85-84
NIOSH Division
DRC;
Source Name
NTIS: PB/84-184209. Available for Reference At Bureau of Mines Libraries.
Performing Organization
Univ. of San Francisco