Skip directly to search Skip directly to A to Z list Skip directly to page options Skip directly to site content

NIOSHTIC-2 Publications Search

Search Results

Failure Report No. S180.

Authors
Anonymous
Source
NIOSH 1979 Apr:28 pages
Link
NIOSHTIC No.
00107295
Abstract
The NIOSH analytical method S180 for determining maleic-anhydride (108316) in air is described. Samples are collected in a midget bubbler, hydrolyzed to maleic-acid (110167), and analyzed by high pressure liquid chromatography with an ultraviolet detector at 254 nanometers. The working range of the method is 0.5 to 2.14 milligrams per cubic meter, the relative standard deviation is 0.63, and the detection limit is 15 micrograms per sample. Analytical apparatus, reagents and procedures are described and calculations are provided for determining sample concentrations. Data on stability, collection efficiency and recovery rates is included. The method was unsatisfactory due to inadequate sample stability.
Keywords
NIOSH-Sampling-Data-Sheet; NIOSH-Backup-Data-Sheet; Air-contamination; Air-sampling; Sampling-methods; Sampling-equipment;
CAS No.
108-31-6; 110-16-7;
Publication Date
19790413
Fiscal Year
1979
NTIS Accession No.
PB81-229650
NTIS Price
A03
Source Name
NIOSH, Unpublished, Cincinnati, Ohio, 28 pages, 8 references
State
OH;