Rotenone Analytical Method.

Authors
Anonymous
Source
NIOSH 1979 May:32 pages
Link
NIOSHTIC No.
00107273
Abstract
The NIOSH analytical method S300 for determining rotenone (83794) in air is described. Samples are collected on a polytetrafluoroethylene membrane filter, extracted with acetonitrile and analyzed by high performance liquid chromatography with an ultraviolet detector at 290 nanometers. The working range of the method is 1.16 to 11.1 milligrams per cubic meter, the coefficient of variance is 0.079, and the detection limit is about 4.0 micrograms. Analytical apparatus, reagents and procedures are described, and calculations are provided for determining sample concentrations. Data on collection efficiency, and recovery and stability rates is included.
Keywords
NIOSH-Analytical-Method; NIOSH-Sampling-Data-Sheet; NIOSH-Backup-Data-Sheet; Analytical-Method; Standards; Insecticides; Analytical-instruments; Standards-Completion-Program; Sampling; SCP; Ultraviolet-radiation;
CAS No.
83-79-4;
Publication Date
19790511
Fiscal Year
1979
NTIS Accession No.
PB82-114729
NTIS Price
A04
Identifying No.
S300
NIOSH Division
DPSE;
Source Name
NIOSH, Unpublished, Cincinnati, Ohio, 32 pages, 8 references