Skip directly to search Skip directly to A to Z list Skip directly to page options Skip directly to site content

NIOSHTIC-2 Publications Search

Search Results

Evaluation of portable, direct-reading combustible gas meters.

Authors
McCammon-CS Jr.
Source
NIOSH 1974 Jun; :1-49
Link
NIOSHTIC No.
00024941
Abstract
Report of a NIOSH test of nine portable gas meters designed to measure combustible gases for 20 specific design and performance characteristics. Known concentrations of hexane, benzene, and methane are dynamically generated in the range of 0-100% Lower Flammable Limit to test the meters' performance. Based on the results of the test, realistic performance specifications are recommended for two classes of combustible gas meters, which will be used for future NIOSH certification regulations for combustible gas meters.
Keywords
Flammable-gases; Aromatic-hydrocarbons; Aliphatic-hydrocarbons; Performance-testing; Measurement-methods; Hazardous-materials; Air-quality-control; Measurement-instrumentation; Calibration
CAS No.
71-43-2; 110-54-3; 74-82-8
Publication Date
19740601
Document Type
Other
Fiscal Year
1974
NTIS Accession No.
PB-232755
NTIS Price
A04
NIOSH Division
DLCD
Source Name
NIOSH Technical Information, Division of Laboratories and Criteria Development, NIOSH, Cincinnati, Ohio, 49 pages, 1 reference
State
OH